The ECG is one of the most useful testing procedures in medicine, used to determine the electrical activity of the heart. It is used for monitoring while other procedures are being performed, as well as for diagnostic purposes.
But an ECG is only as accurate as the positioning of its leads and the ability of the technician administering it to observe proper protocols.
This short guide will cover ECG lead wire positioning and importance.
The Importance of Proper ECG Lead Wire Positioning
While preparation of the patient’s skin and the use of high-quality ECG lead wires is critical to the accuracy of an ECG recording, the placement of the electrodes is equally as important.
Improper ECG lead wire placement can result in inaccurate recordings which can have serious consequences for patient diagnoses. Poor lead wire positioning can result in misinterpretations and erroneous interpretations of an ECG.
Therefore, proper lead wire positioning is fundamental to accuracy.
3-Lead ECG Positioning
In a 3-lead ECG, only three electrodes and lead wires are used. These are the right arm (RA), left arm (LA), and left leg (LL) leads.
For the right arm and left arm leads, the electrodes should be placed between the shoulder and the elbow of each arm, respectively. The left leg lead should be placed below the left torso and above the left ankle.
Some guides suggest that for best results, the RA and LA leads should be placed on the chest wall equidistant from the heart, rather than on the indicated limbs themselves; this helps to reduce motion artifacts.
5-Lead ECG Positioning
In the 5-lead ECG, the RA, LA, and LL leads are also used, along with right leg (RL) and a chest lead (called V1 or C).
The positioning of the RA and LA leads is the same. For the right leg lead, the electrode is usually placed over the edge of the rib cage in the 7-8 intercostal space, at the midclavicular line. The left leg lead is placed at the edge of the rib cage on the other side, in the 7-8 intercostal space at the midclavicular line.
Placing the RA, LA, RL, and LL leads on the chest instead of on the arms and legs helps to reduce motion artifacts.
The last lead, the chest lead, is placed in the 4th intercostal space, just to the right side of the sternum.
12-Lead ECG Positioning
Despite the name, the 12-lead ECG uses 10 electrodes and ECG lead wires. In the 12-lead ECG, the positioning of the RA, LA, LL, and RL leads is the same.
There are also 6 additional precordial leads, the V1-V6 leads. These are positioned as follows:
● V1: The 4th intercostal space, to the right of the sternum (this is analogous to the single chest lead used in a 5 lead ECG).
● V2: The 4th intercostal space, at the left margin of the sternum.
● V3 (Position V4 first, see below): Halfway between V2 and V4.
● V4: the 5th intercostal space, at the midclavicular line.
● V5: The 5th intercostal space, at the anterior axillary line.
● V6: The 5th intercostal space, at the mid-axillary line.
High-Quality ECG Lead Wires
The use of high-quality ECG lead wires, like proper positioning, is also necessary to produce an accurate recording.